Gas Chromatography 30-GCG100

Gas Chromatography 30-GCG100


Gas chromatography system comes with different detectors and each detector individually controls the temperature of the system. This gas chromatography comes with packed column injectors, split/split less injectors, capillary and cool-on column. It is possible to achieve lower limit of detection using these unique injector system. It gives accurate, high precision control, reliable results.



Chromatography column typeCapillary and Packed column
Chromatography column length100 to 500 mm
Chromatography column diameter0.2 to 0.5 mm
Temperature operating rangeHigher than the room temperature of 5 ℃ to 450 ℃
Maximum programmed heating speed120 ℃/min
Programmed temperature steps7 steps
Temperature setting precision1 °C
Programmed heating rate0.1 to 120 °C/min (increment of 0.1 °C)
Furnace dimension280 × 300 ×180 mm
InjectorsPacked column injectors, split/split less capillary column injectors, cool on-column injectors
DetectorsFollowing detectors can be selected as per the requirement
FIDMaximum temperature control: 400 °C
LOD: ≤ 5 × 1012 g / s [n - C16]
Linear dynamic range: ≥ 107
Drift: ≤ 5 × 10-13 A / 30 min
Noise: ≤ 2 × 10-13 A
TCDMaximum temperature control: 400 °C
LOD: ≥ 10000 mv.ml / mg [n - C16]
Linear dynamic range: ≥ 104
Drift: ≤ 10 µV / 30 min
Noise: ≤ 20 µA
FPDMaximum temperature control: 250 °C
LOD: ≤2×10-13 g / s (P), ≤ 5 × 10-11 g / s (S)
Linear dynamic range: ≥ 103 (P), ≥ 102 (S)
Drift: ≤ 4 × 10-11 A / 30 min
Noise: ≤ 2 × 10-12 A
ECDMaximum temperature control: 350 °C
LOD: ≤ 1 × 10-14 g / ml (y - 666)
Linear dynamic range: ≥ 104
Drift: ≤ 5 µV / 30 min
Noise: ≤ 20 µA
NPDMaximum temperature control: 400 °C
LOD: ≤ 5 × 10-13 g / s (P), ≤ 1 × 1012 g / s (N)
Linear dynamic range: ≥ 103 (P), ≥ 103 (N)
Drift: ≤ 2.5 × 10-12A / 30 min
Noise: ≤ 4 × 10-13 A
PDHIDLOD: ≤ 5 ppb
  • 7 steps temperature rising function
  • LCD display to view test parameters
  • Individual temperature controlling detectors
  • Self-diagnosis, keyboard locking function
  • Data storage during power off
  • Different detectors with high stability and sensitivity for selection
  • Stable and speedy program for furnace heating
  • Unique injectors system for detection of lower limit
Used in the analysis of oil extraction and refining industry, petrochemical analysis, gas industry and in quality control of medicines, food and chemicals.