Gas chromatography mass spectrometry is a high precision Gas chromatograph mass spectrometer with pre-filter mass analyzer and electron multiplier ensuring high sensitivity. Desorption for solids is performed at controlled temperature between 100 ~ 350 °C with a flow rate of 1 to 1.5 ml / min. With seven stage temperature programming and an interface temperature of 450 °C the analysis is sensitive and specific.

Gas Chromatography Specifications a:12:{s:8:"Model No";s:5:"GC-87";s:9:"Flow rate";s:16:"1 ~ 1.5 ml / min";s:17:"Maximum flow rate";s:11:"10 ml / min";s:17:"Inlet temperature";s:7:"450 °C";s:14:"Pressure range";s:11:"0 ~ 100 psi";s:12:"Heating rate";s:18:"Upto 120 °C / min";s:16:"Room temperature";s:17:"4 °C ~ 450 °C";s:21:"Pressure control mode";s:50:"Electronic pressure control ( Supports CV and CC )";s:10:"Split mode";s:18:"Split / split less";s:11:"Split ratio";s:8:"1000 : 1";s:23:"Temperature programming";s:22:"7 stages / 8 platforms";s:12:"Auto sampler";s:8:"Optional";}
Mass Spectrometry Specifications a:12:{s:8:"Model No";s:5:"MS-89";s:10:"Mass range";s:14:"1.5 ~ 1000 amu";s:26:"Ion – source temperature";s:19:"100 °C ~ 350 °C";s:27:"GC-MS interface temperature";s:7:"450 °C";s:9:"Scan rate";s:16:"Upto 10000 amu/s";s:9:"Stability";s:21:"± 0.10 amu / 48 hrs.";s:25:"Filament emission current";s:11:"0 ~ 350 µA";s:27:"EI source ionization energy";s:13:"5 eV ~ 250 eV";s:10:"Resolution";s:15:"Unit resolution";s:8:"Detector";s:38:"High energy dynode electron multiplier";s:11:"Sensitivity";s:33:"Full scan ( S / N is ≥ 30 : 1 )";s:6:"Vacuum";s:31:"Turbo molecular pump ( 67 L/s )";}
  • Column flow rate is 1 ~ 1.5 ml / min
    Maximum flow rate is 10 ml / min
    Simultaneous analysis with different components increasing productivity
    Electronic flow control system
    Mass analyzer with pre-filter reducing quadruple filtration
    Turbo molecular pump vacuum system for stability and reliability
    Software controlled auto sampler , gas chromatograph and mass spectrometer
    Screens and quantifies more than 1000 compounds in a single run
  • Used in identifying and quantifying volatile organic compounds in mixtures, testing residual solvents, identification of trace impurities in liquids or gases, evaluating extracts from plastics, contaminants on semiconductor wafers or other technology products

Gas Chromatography

Gas Chromatography 30-GCG100

  • Chromatography column type : Capillary and Packed column
  • Chromatography column length : 100 to 500 mm

Gas Chromatography 30-GCG101

  • Chromatography column type : Capillary or Packed column
  • Chromatography column length : 100 to 300 mm

Gas Chromatograph 65-GC104

  • Temperature Range : RT to 450 °C
  • Temperature Accuracy : ≤ ± 0.1℃

Ion Chromatography

Ion chromatograph system 30-ICG200

  • Ion Chromatographic Pump : a:6:{s:4:"Type";s:64:"High pressure and low pulse double plunger tandem advection pump";s:16:"Maximum pressure";s:13:"35 MPa (PEEK)";s:25:"Pressure display accuracy";s:11:"≤ 0.1 MPa";s:14:"Flow stability";s:91:"(0.2 to 0.5) ml/ min ≤ 3 %
    (0.5 to 1.0) ml/ min ≤ 2 %
    > 1.0 ml/ min 2 %
    ";s:10:"Flow range";s:22:"0.001 to 9.999 ml/ min";s:14:"Pressure pulse";s:8:"≤ 0.5%";}
  • Conductivity detector : a:16:{s:4:"Type";s:53:"Temperature Control and Bipolar Conductivity Detector";s:14:"Detection mode";s:30:"Bipolar conductivity detection";s:11:"Cell volume";s:11:"≤ 0.8 µL";s:15:"Detection range";s:18:"0 to 35000 µS/ cm";s:20:"Detection resolution";s:17:"≤ 0.0020 nS/ cm";s:14:"Output voltage";s:31:"- 6000 to +6000 mV (adjustable)";s:14:"Baseline noise";s:15:"≤ 0.05 % F.S.";s:14:"Baseline drift";s:18:"≤ 3 % FS/ 30 min";s:27:"Operating temperature range";s:43:"Room temperature +5 to 60°C (41 to 140°F)";s:32:"Controlling temperature accuracy";s:10:"± 0.01°C";s:16:"Maximum pressure";s:6:"10 MPa";s:12:"Linear range";s:7:"≥ 103";s:21:"Instruments linearity";s:9:"≥ 0.999";s:26:"Quantitative repeatability";s:9:"≤ 1.0 %";s:25:"Qualitative repeatability";s:9:"≤ 1.0 %";s:32:"Minimum detectable concentration";s:73:"Cl ≤ 0.0005 ug/ ml
    BrO3 ≤ 0.001 ug/ ml
    Li+ ≤ 0.001 ug/ ml";}

Ion chromatograph system 30-ICG201

  • Ion Chromatographic Pump : a:6:{s:4:"Type";s:64:"High pressure and low pulse double plunger tandem advection pump";s:16:"Maximum pressure";s:13:"35 MPa (PEEK)";s:25:"Pressure display accuracy";s:11:"≤ 0.1 MPa";s:14:"Flow stability";s:86:"(0.2 to 0.5) ml/ min ≤ 3 %
    (0.5 to 1.0) ml/ min ≤ 2 %
    > 1.0 ml/ min 2 %";s:10:"Flow range";s:22:"0.001 to 9.999 ml/ min";s:14:"Pressure pulse";s:8:"≤ 0.5%";}
  • Conductivity detector : a:16:{s:4:"Type";s:53:"Temperature Control and Bipolar Conductivity Detector";s:14:"Detection mode";s:30:"Bipolar conductivity detection";s:11:"Cell volume";s:11:"≤ 0.8 µL";s:15:"Detection range";s:18:"0 to 45000 µS/ cm";s:20:"Detection resolution";s:17:"≤ 0.0020 nS/ cm";s:14:"Output voltage";s:31:"- 6000 to +6000 mV (adjustable)";s:14:"Baseline noise";s:15:"≤ 0.05 % F.S.";s:14:"Baseline drift";s:18:"≤ 3 % FS/ 30 min";s:27:"Operating temperature range";s:43:"Room temperature +5 to 60°C (41 to 140°F)";s:32:"Controlling temperature accuracy";s:10:"± 0.01°C";s:16:"Maximum pressure";s:6:"10 MPa";s:12:"Linear range";s:7:"≥ 103";s:21:"Instruments linearity";s:9:"≥ 0.999";s:26:"Quantitative repeatability";s:9:"≤ 1.0 %";s:25:"Qualitative repeatability";s:9:"≤ 1.0 %";s:32:"Minimum detectable concentration";s:73:"Cl ≤ 0.0005 ug/ ml
    BrO3 ≤ 0.001 ug/ ml
    Li+ ≤ 0.001 ug/ ml";}