Preparative HPLC 37-PHP100

Preparative HPLC 37-PHP100

Preparative HPLC is equipped with 4 pump head variation and a UV or RI detector with a wavelength range of 190 to 400 nm and 400 to 800 nm. Built in multi positioned fraction valve and controlled fractionation optimizes controlled fraction collection and solvent recycling. With a column size of 10 x 1000 mm the retention, resolution and selectivity is optimized.

Dimension 750 x 860 x 1280 mm ( 100 ml , 250 ml , 500 ml , 1000 ml ) , 810 x 950 x 1310 mm ( 3 L )
Wetted material 316 L and PTFE
Power < 1500 W
Mobility With castor
  • Solvent flow rate is 500 ml / min
    Multiple chromatogram view
    27 integrated programmable parameters like peak width, tangent slope etc
    Compatibility for medium and low pressure application
    10 points pressure calibration for pressure linearity
    Auto apply of mathematical operations to chromatograms
    Automated sequential commands post batch run
    External pump control by panel or computer software
    Addition of tools for detector controlled fraction collection and solvent recycling
    Controlled fractionation with option of manual control during the run
    Error alarm function for over pressure ,under pressure and leakage
  • Used for purification of peptides and proteins, separation of high molecular mass compounds, separation of protein mixtures, qualitative and quantitative determination of a compound, automated fraction re-analysis

Analytical HPLC

Analytical HPLC 37-AHP100

  • Pump Specifications :
    Model No HPP-A1
    Flow rate range 0.0001 ~ 10 ml / min
    Maximum pressure 50 MPa , 0 ~ 2 ml / min
    Column size 4.6 x 250 mm11
    Sample size 5 µm
    Pressure pulsation ≤ 1 % at 10 MPa
    Flow rate accuracy ± 0.15 % at 1 ml / min
    Flow rate repeatability % ≤ 0.1
    Gradient accuracy % ± 1.5
    Gradient repeatability % ≤ 0.2
    Dimension 460 x 310 x 180 mm
    Power supply 90 V
    Weight 11.6 kg
  • Detector Specifications :
    Model No HPD-A20 HPD-A21
    Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm )
    Wavelength accuracy ± 0.5 nm ± 2 nm
    Wavelength repeatability 0.15 nm 0.3 nm
    Band width 8 nm 8 nm
    Optical path 10 nm 10 nm
    Deuterium lamp 190 ~ 400 nm 190 ~ 400 nm
    Tungsten lamp 400 ~ 800 nm 400 ~ 800 nm
    Baseline noise ( Static ) Au ≤ ± 0.35 x 10 -5 ≤ ± 0.5 x 10 -5
    Baseline drift ( Static ) Au / h ≤ 1 x 10 -4 ≤ 1.5 x 10 -4
    Minimum detection limit ≤ 4 x 10 -8 ≤ 4 x 10 -8
    Power 100 W 100 W
    Power supply 90 V / 50 ~ 60 Hz 90 V / 50 ~ 60 Hz
    Dimension 460 x 315 x 180 mm 460 x 315 x 180 mm
    Weight 9.5 kg 9.5 kg

Analytical HPLC 37-AHP101

  • Pump Specifications :
    Model No HPP-B1
    Flow rate range 0.0001 ~ 10 ml / min
    Maximum pressure 42 MPa , 0 ~ 5 ml / min
    Column size 4.6 x 250 mm11
    Sample size 5 µm
    Pressure pulsation ≤ 1.5 % at 10 MPa
    Flow rate accuracy ± 0.2 % at 1 ml / min
    Flow rate repeatability % ≤ 0.15
    Gradient accuracy % ± 2
    Gradient repeatability % ≤ 0.2
    Dimension 450 x 260 x 190 mm
    Power supply 90 V ~ 240 V
    Weight 11.2 kg
  • Detector Specifications :
    Model No HPD-B2
    Wavelength range 190 to 800 nm
    Wavelength accuracy ± 0.75 nm
    Wavelength repeatability 0.2nm
    Band width 8 nm
    Optical path 10 nm
    Deuterium lamp 190 ~ 400 nm
    Tungsten lamp 400 ~ 800 nm
    Baseline noise ( Static ) Au ≤ ± 0.5 x 10 -5
    Baseline drift ( Static ) Au / h ≤ 1.5 x 10 -4
    Minimum detection limit ≤ 4 x 10 -8
    Power 100 W
    Power supply 90 V / 50 ~ 60 Hz
    Dimension 460 x 260 x 190 mm
    Weight 9.2 kg

Semi-prep HPLC

Semi-prep HPLC 37-SPH100

  • Pump Specifications :
    Model No SPH-A10 SPH-A11 SPH-A12 SPH-A13
    Flow rate range 0.001 ~ 50 ml / min 0.001 ~ 50 ml / min 0.001 ~ 100 ml / min 0.001 ~ 100 ml / min
    Maximum pressure 30 MPa 30 MPa 15 MPa 10 MPa
    Flow rate accuracy ± 0.35 ± 0.5 ± 0.5 ± 0.5
    Flow rate repeatability % ≤ 0.1 ≤ 0.15 ≤ 0.1 ≤ 0.15
    Column size 10 x 250 mm 10 x 250 mm 10 x 250 mm 10 x 250 mm
    Sample loading 35 ~ 60 mg 35 ~ 60 mg 35 ~ 60 mg 35 ~ 60 mg
    Pressure pulsation 1 % at 10 MPa 1.5 % at 10 MPa 1.5 % at 10 MPa 1.5 % at 10 MPa
    Flow rate repeatability % ≤ 0.1 ≤ 0.15
    Gradient accuracy % ± 2 ± 2 ± 2 ± 2
    Gradient repeatability % ≤ 0.2 ≤ 0.2 ≤ 0.2 ≤ 0.2
    Dimension 465 x 315 x 180 mm 455 x 265 x 190 mm 465 x 315 x 180 mm 465 x 315 x 180 mm
    Power 150 W 150 W 150 W 150 W
    Power supply 90 V ~ 240 V/ 50 Hz 90 V ~ 240 V/ 60 Hz 90 V ~ 240 V/ 50 Hz 90 V ~ 240 V/ 60 Hz
    Weight 11.6 kg 11.2 kg 11.6 kg 11.6 kg
  • Detector Specifications :
    Model No SPH-A20 SPH-A21 SPH-A22
    Wavelength range 190 to 800 nm 190 to 800 nm 190 to 800 nm
    Wavelength accuracy ± 0.5 nm ± 1 nm ± 0.75 nm
    Wavelength repeatability 0.15 nm 0.3 nm 0.2 nm
    Band width 8 nm 8 nm 8 nm
    Optical path Semi flow / analytical Semi flow / analytical Semi flow / analytical
    Baseline noise ( Static ) Au ≤ ± 0.35 x 10 -5 ≤ ± 0.5 x 10 -5 ≤ ± 0.5 x 10 -5
    Baseline drift ( Static ) Au / h ≤ 1 x 10 -4 ≤ 1.5 x 10 -4 ≤ 1.5 x 10 -4
    Minimum detection limit ≤ 4 x 10 -8 ≤ 4 x 10 -8 ≤ 4 x 10 -8
    Power 150 W 150 W 150 W
    Power supply 90 V / 50 ~ 60 Hz 90 V / 50 ~ 60 Hz 90 V / 50 ~ 60 Hz
    Dimension 465 x 315 x 180 mm 465 x 315 x 180 mm 460 x 265 x 190 mm
    Weight 9.5 kg 9.5 kg 9.2 kg

Standard HPLC

Enquiry Form
Captcha Image